ISO 9001 equipment calibration record

If a piece of equipment is used to decide whether a product conforms, the company must be able to show that it measures correctly: when it was calibrated, against what, with what result and what was done when it was out.

In one sentence

The equipment calibration record shows that the equipment used to verify product conformity is fit for purpose: ISO 9001 requires evidence of that fitness to be retained (7.1.5.1) and, where measurement traceability is needed, the equipment to be calibrated or verified at specified intervals against traceable standards, its status identified and previous results reviewed if a device is found to be unfit (7.1.5.2).

Content reviewed on

Where the standard requires it
ISO 9001 · Clause 7.1.5 Clauses are cited using the numbering of the 2015 edition, the same one the Academy uses; certificates issued against it remain valid during the transition to the 2026 edition.
Type of information
Mandatory record Documented information to be retained (7.1.5.1)
Who fills it in
Quality or maintenance External calibration is done by a laboratory; the decision on the result is made internally
When
At the interval set for each device And whenever the device is knocked, repaired or gives a doubtful result
How long it is kept
Set by the organisation The standard sets no periods: it asks you to decide on retention and disposition (7.5.3.2). Covering at least the three-year certification cycle lets you show the history at recertification.

What the standard requires


What ISO 9001 demands and what it leaves to your judgement

Where monitoring or measuring is used to verify that products and services conform, the organisation must have suitable resources, keep them fit for purpose and retain evidence of that fitness.

Where measurement traceability is a requirement, or the company considers it essential, equipment is calibrated or verified at specified intervals or before use, against standards traceable to national or international measurement standards; where no such standard exists, the basis used is retained. Equipment must be identified so its status is known, and protected from adjustment, damage or deterioration.

If a device is found to be unfit, you must determine whether the validity of previous measurement results has been compromised and act accordingly. This is the part most often forgotten.

Structure


Fields the template should include

  • Device, internal code, location and use
  • Range and tolerance or maximum permissible error for that use
  • Type of control: external calibration or internal verification
  • Interval and date of next calibration
  • Reference standard or laboratory and its traceability
  • Result: error found and uncertainty
  • Decision: fit, fit with restrictions, unfit
  • If unfit: analysis of previous measurements and actions
  • Status identification on the device itself

Example


Example: micrometer found out of tolerance

Company, people and figures are fictitious: the example only shows what a properly completed record looks like.

Field Content
Device Outside micrometer 0-25 mm · MIC-07 · final inspection
Tolerance for use Maximum permissible error ± 0.004 mm
Control Annual external calibration at an accredited laboratory
Result Error of + 0.007 mm at 20 mm · dated 9 March 2026
Decision Unfit: withdrawn and sent for adjustment
Previous measurements 14 batches measured since the previous calibration reviewed; 2 with dimensions close to the upper limit
Action Remeasurement of the retained samples from the 2 batches: conforming · customer not affected
Status "Out of use" label until its return; new certificate on 23 March

At the audit


Mistakes auditors commonly flag

  • Calibration certificates filed without anyone comparing the result with the tolerance for use.
  • Equipment out of tolerance with no analysis of what has been measured with it since the previous calibration.
  • Measuring equipment in use that does not appear in the inventory.
  • Calibration intervals set without any rationale and never reviewed.
  • Overdue equipment not identified on the line.

In Kimobox


How this record is kept in Kimobox

  1. 1Each measuring device has its own record with location, use and full history.
  2. 2Calibrations and verifications are scheduled at their intervals, and the system warns you before each one falls due.
  3. 3The certificate and the fitness decision are kept in the device's history, linked to the quality system.
  4. 4An unfit device is handled as one more nonconformity, with cause, action and verification of effectiveness.

Frequently asked questions


What people ask us before getting started

Does all equipment have to be calibrated?

No. The standard refers to monitoring and measuring resources used to verify the conformity of products and services. A room thermometer in an office is not included; the calliper used to release a batch is.

Are calibration and verification the same thing?

No. Calibrating means comparing against a reference standard and knowing the device's error; verifying means checking that this error is within what the use allows. ISO 9001 accepts one, the other or both, depending on the case.

What calibration interval should be set?

The standard does not set one. It is decided according to use, the device's stability, the manufacturer's recommendations and the history of results, and it should be reviewed: if a device always comes out comfortably within tolerance, the interval can be extended with justification.

Keep your ISO 9001 records without loose templates

We'll show you the module running with data similar to yours.

Request a demo